bpAcc: Blood Pressure Device Accuracy Evaluation: Statistical Considerations

A comprehensive statistical analysis of the accuracy of blood pressure devices based on the method of AAMI/ANSI SP10 standards developed by the AAMI Sphygmomanometer Committee for indirect measurement of blood pressure, incorporated into IS0 81060-2. The 'bpAcc' package gives the exact probability 'of accepting a device D' derived from the join distribution of the sample standard deviation and a non-linear transformation of the sample mean for a specified sample size introduced by Chandel et al. (2023) and by the Association for the Advancement of Medical Instrumentation (2003, ISBN:1-57020-183-8).

Version: 0.0-2
Depends: R (≥ 4.2.0), methods, stats
Published: 2024-01-18
DOI: 10.32614/CRAN.package.bpAcc
Author: Victor Miranda [cre, aut, cph], Tanvi Chandel [aut], Andrew Lowe [aut], Tet Chuan Lee [aut], Auckland_University_of_Technology (AUT) [cph, fnd]
Maintainer: Victor Miranda <victor.miranda at aut.ac.nz>
License: GPL-2
NeedsCompilation: yes
Materials: NEWS
CRAN checks: bpAcc results

Documentation:

Reference manual: bpAcc.pdf

Downloads:

Package source: bpAcc_0.0-2.tar.gz
Windows binaries: r-devel: bpAcc_0.0-2.zip, r-release: bpAcc_0.0-2.zip, r-oldrel: bpAcc_0.0-2.zip
macOS binaries: r-release (arm64): bpAcc_0.0-2.tgz, r-oldrel (arm64): bpAcc_0.0-2.tgz, r-release (x86_64): bpAcc_0.0-2.tgz, r-oldrel (x86_64): bpAcc_0.0-2.tgz
Old sources: bpAcc archive

Linking:

Please use the canonical form https://CRAN.R-project.org/package=bpAcc to link to this page.